Archwire (Composite wire) MRI Safe
Archwire (NiTi wire) Suggested to remove because of artifacts
Body Piercings (and Jewelry) Metallic body piercing and jewelry should be removed prior to entering the MR environment
No piercings (jewelries) in the area to be examined (imaging area)
Piercings and jewelries made of plastic (Teflon or BioFlex) are OK
If it is not possible to remove metallic body piercing (jewelry) and it is not in the imaging area, investigate its composition
The patient or individual should be informed regarding the potential risks, the piercing should be stabilized (e.g., application of adhesive tape or bandage) and wrapped
Prior to the MRI exam, the patient should be instructed to immediately inform (through the panic ball) the L2 if heating or other unusual sensation occurs in association with the body piercing (jewelry)
Take breaks (3 minutes) between high SAR sequences (EPI/DTI)
Brackets Dental (Ceramic or plastic) MRI Safe
Brackets Dental (Ceramic with SS slot) Should be removed because of artifacts in the oral cavity
Brackets Dental (Titanium) Should be removed because of artifacts in the oral cavity

Contact Lenses
Intraocular Lens (IOL) Implants

Standard contact lenses do not usually pose a risk for the MRI examination but it is strongly advised to remove them
IOL do not usually pose a risk for the MRI examination but it is strongly advised to remove them because in a few older models, metallic loops made of platinium, iridium or titanium may be attached to the lens for suturing purposes

Exceptions:
Color Contact Lenses are NOT allowed and must be removed
Sensimed Triggerfish Contact Lenses are NOT allowed in the MR environment

Colonoscopy (in the last eight weeks) Check whether any endoscopy clips or pill cam were used or placed during the procedure with the patient or through patient's medical records
If existence is confirmed or unknown you are NOT allowed to scan 
Possible Artifacts
Patients must be encouraged to squeeze the communication ball if they feel any warmth or strange sensation
Dental bone screws Remove it if possible
If it is not possible make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)
Dental Crowns: Amalgam, gold alloy, gold ceramic, titanium and titanium alloy, NiTi, some composites, ceramic, glass ionomers, resin, gutta percha, zirconium dioxide Maybe artifacts in the oral cavity
The patient or individual should be informed regarding the potential risks (dislodging sensations or heating)
Prior to the MRI exam, the patient should be instructed to immediately inform (through the panic ball) the L2 if any unusual sensation occurs
Dental Crowns: ferromagnetic components (Stainless steel) Make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR <3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)
Dental Implants Remove it if possible
If it is not possible make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)
EEG Electrodes Only those that are MR Safe
ECG or EKG Electrodes Only those that are MR Safe or those provided by Siemens
Eye Prostheses

They should be MR Safe

Make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)

Eyelid Weights and Springs Remove them if possible
If it is not possible make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)
Glaucoma Devices Remove it if possible
If it is not possible make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)
Hip/Knee Replacements Must be at least 6 weeks post op
Make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)
IUDs Make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php  
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)
Unknown IUDs are not accepted
Ligature Wire (Stainless Steel) Suggested to remove because of artifacts
Makeup (glittery) May contain metallic particles. Investigate
Maxillofacial prostheses 

Make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)

Nail polish

May contain metallic particles. Investigate

Permanent Cosmetics Once the presence of the permanent cosmetics has been identified (tattooed eyeliner, eyebrows, lips, cheeks and nipple-areola reconstruction) you must investigate the pigments
Pigment colors (e.g., brown, black, red, and flesh) typically contain iron oxide
A cold compress (e.g., ice bag) or cold wet towel may be applied to the tattoo site
Remember that permanent eyeliner and periorbital tattoos may cause local swelling and pain even hours after scanning
The patient should be informed of the risks associated with the site of the tattoo
The patient should be advised to immediately inform the L2 (through the panic ball) regarding any unusual sensation felt at the site of the tattoo in association with the MR procedure [1]
Take breaks (3 minutes) between high SAR sequences (EPI/DTI)
Retainers Dental (Fixed) containing steel Should be removed because of artifacts in the oral cavity
Retainers Dental (Fixed) with fiber wire MRI Safe
Scleral Banding (Buckling) Make and model merit investigation
Search into the website of the device manufacturer (up to which field)
Try to find information in http://www.mrisafety.com/TMDL_list.php 
Only Normal Operating Mode allowed in 3T Field
Max Whole Body SAR < 2W/Kg and Max Head SAR < 3.2 W/Kg
Highest Spatial Gradient Magnetic Field for the MR Conditional implant/device rated as acceptable up to a value of 7 T/m (700 Gauss/cm)
Tattoos Tattoos made before the year 2000 are strictly NOT allowed
Allowed Tattoos are only those below the sternal angle and older than six months
No Tattoos > 20 cm length
The patient should be informed of the risks associated with the site of the tattoo
The patient should be advised to immediately inform the L2 (through the panic ball) regarding any unusual sensation felt at the site of the tattoo in association with the MR procedure
Take breaks (3 minutes) between high SAR sequences (EPI/DTI)
Transdermal or Medication Patches Allowed only if you can remove them
Never remove a medication patch without referring MD approval [2]
 

 

References
[1] https://onlinelibrary.wiley.com/doi/full/10.1002/jmri.10049 
[2] https://www.ismrm.org/mr-safety-links/mr-safety-week-2017/mri-safety-and-the-story-of-a-medication-patch/